TDCS Neurostim

Description

Transcranial Electrical Stimulator


  • DC output current – up to 2mA
  • Continuous monitoring of electrode impedance
  • Two modes of use: inpatient and outpatient
  • Storage – up to 250 sessions
  • USB connection for advanced control

Exellence

Stimulus current - up to 5mA (30V)
Stimulus duration - up to 40 min
Stimulus Type ;
tDCS
tACS
tRNS
tDCS-Sham
tPCS

Specification

Number of channels                                                              : 1
Stimulation Channel    
Pulse type : Current : Current
Impedance measurement error (patient circuit) within the range 0.5 to 12.0 kΩ at stimulation current
0.3 mA to 5 mA current                                                           
: ± (10% of the specified value +200 Ω)
Maximum voltage at channel output (without load)         : not more than 38 V
Pulse amplitude                                                        : 0 to 2 mA
Adjustment step for pulse amplitude                      : 25 µA
Session duration                                           : 15 s to 40 min
Session duration adjustment step                                         : 1 s
Stimulation mode/pulse waveform:    
• tDCS : transcranial direct current stimulation with adjustable amplitude, ramp up/down time, and sham  stimulation option
• tPCS                                                                                         : transcranial pulsed current stimulation by monophasic square pulse waveform with adjustable amplitude, duration, and interpulse interval
• tACS : transcranial alternating current stimulation by sinusoidal, bipolar square or custom (user-defined) pulse waveform with adjustable amplitude, offset, and start phase
• tRNS : transcranial random noise stimulation by noise-like pulses with adjustable amplitude and offset
Direct current in patient circuit at disabled stimulation output  : not more than 10 µA
Sampling period of stimulation current:
• tDCS                                                                                          
: 2 ms
• tACS                                                                                          : 30 µs
• tRNS : 100 µs
Stimulation Mode Parameters    
tDCS — transcranial direct current stimulation    
Current ramp up/down time                                                    : 0 to 2 min
Adjustment step for current ramp up/down time               : 1 s
Current ramp up/down time during sham stimulation       : 0 to 2 min
Adjustment step for current ramp up/down time during sham stimulation  : 1 s
tPCS — transcranial pulsed current stimulation    
Frequency                                                                                   : 1 to 250 Hz
Frequency adjustment step: :  
• within the range from 1.00 to 10.00 Hz         : 0.01 Hz
• within the range from 10.0 to 250.0 Hz                             : 0.1 Hz
Offset                                                                                        : – 1000 to 1000 µA
Offset adjustment step                                                          : 10 µA
Start phase Start                                                                        : 0 to 355°
phase adjustment step                                                             :
Pulse waveform                                                                         : sinusoidal, square, custom
tRNS — transcranial random noise stimulation :  
Offset                                                                                            : – 1000 to +1000 µA
Offset adjustment step                                                       : 10 µA
General Specifications    
Connection to PC                                                                      : USB 2.0
Maximum number of sessions                                             : 250
Power supply    
one 6LR61 battery (nominal voltage – 9 V, nominal capacity – not less than 550 mAh)    
(5±0.25) V, from 300 mA (PC)    
Power consumption                                                                  : not more than 2.5 W
Personal Computer Requirements    
Operating system :
• Windows 10×64/×86
• RAM: 2 GB and more
• free disk space (hard drive): not less than 50 MB
• Intel Core i3-6300 or AMD Ryzen 3 3200G processor and more powerful
• USB 2.0
   

Standard Configurations

Main Unit tDCS : 1
USB Cable : 1
Cable for electrode Connection : 2
Conductive Silicone electrode : 2
Spontex holding bag : 2
Electrode Fixation : 2
Rubber Tape Fixator : 2
Manual Book : 1
Software Neurostim : 1
Laptop : 1