Assessment of central and peripheral motor pathways
Magnetic stimulation has proved effective in the assessment of central and peripheral motor pathways. The alternating magnetic field generated by the stimulator easily passes through clothes, skin, scalp, bones, and meninges. Reaching the conductive tissues of central and peripheral nervous systems, this field creates an alternating electric field which in turn induces an electric current that can activate the neurons as during electrical stimulation. However, if compared to electrical stimulation, magnetic stimulation does not cause painful sensations that usually occur due to the activation of skin receptors under the electrode and does not require additional preparation time. Using an EMG system during stimulation allows recording motor evoked potentials (MEPs) from the muscles innervated by those neurons. This principle underlies a variety of diagnostic tests such as central motor conduction time, cortical silent period, and others.
Diagnostics of neurological disorders
Neuro-MSX magnetic stimulator allows for the quick and painless study of central and peripheral motor pathways which is important for early diagnostics, assessment, prognosis, and monitoring of various neurological disorders including multiple sclerosis, amyotrophic lateral sclerosis, movement disorders, motor neurone disease, and spinal cord injuries.
Compatible with most of EMG systems
Neuro-MSX is compatible with most EMG systems of world manufacturers. If you already have an EMG system, you can use it with our magnetic stimulator. However, the best diagnostic results can be achieved using Neurosoft EMG systems.
Neuro-MEP.NET software supplied with the stimulator features built-in algorithms for a wide range of TMS studies:
|
|
|
|
|
|
All these tests are available when purchasing Neuro-MSX magnetic stimulator with any of our EMG systems.
Wide choice of coils
Depending on the type of study, you can select the coil of any shape and size from a great variety that we have designed. All coils feature perfect ergonomics and controls right at your fingertips while the coordinate grid helps you achieve maximum positioning accuracy.
|
|
|
|
rTMS
| Stimulation Parameters | ||
| Pulse waveform | : | biphasic |
| Pulse duration range | : | 240 … 410 µs |
| Peak magnetic field | : | 2.5 T |
| Rated energy store on the main unit capacitor at the set amplitude of 100 % | : | 320 J |
| Stimulation Mode | : | Single Pulse |
| : | Train | |
| : | Burt | |
| : | Ramp | |
| : | Sweep Frequency | |
| Maximum frequency | : | 100 Hz |
| Maximum pulse frequency in burst | : | 2000 Hz |
| Coil surface temperature enabling automatic stimulation stop | : | (39 ± 2) °C |
| Power Supply | ||
| Input voltage | : | (220 ± 22)/(230 ± 23)/(240 ± 24) V |
| Input frequency | : | 50 or 60 Hz |
| Power consumption during the stimulation | : | |
| Main unit | : | ≤ 3000 VA |
| Cooling unit | : | ≤ 190 VA |
| Extra Power Supply unit | : | ≤ 3000 VA |
| Wireless Connection with PC | ||
| Interface | : | Wi-fi |
| : | IEEE 802,11 b/g/n standard |
qEEG
| Number of Channel | : | 47 channels |
| Sampling Rate | : | 2000 Hz |
| Voltage Range | : | 1 - 12000 μV |
| Ratio error Voltage Measurement | ||
| in the range from 10 up to 50 μV | : | within ±25% |
| in the range from 50 up to 10000 μV | : | within ±7% |
| Sensitivity | : | 0.01 - 10.000.000 μV/mm |
| Ratio error of sensitivity | : | within ±5% |
| High pass filter | : | arbitary value can be set in 0.01 - 10 Hz range |
| Low pass filter | : | arbitary value can be set in 0.1 - 500 Hz range |
| Sweep Speed | : | arbitary value can be set in 1 - 10000 mm/s range |
| Ratio error sweep speed | : | within ±2% |
| Bandpass flatness in 0.5 - 60 Hz | : | from -10 up to +5% |
| Suppresion ratio of power frequency by notch filter | : | not less than 40 dB |
| CMRR | : | not less than 120 dB |
| Input noise level in 0.5 - 200 Hz range (rms) | : | not more than 1.4 μV |
| Input Impedance | : | not less than 400 MΩ |
| Patient leakage current | : | not more than 50 nA |
| Voltage measurement range | : | 0,2 - 100 mV |
Neuronavigation System
| 3-Dimensional Optical Tracking System including: | ||||
| Track the 3D position orientation of active or passive markers attached to the TMS Coil | ||||
| Overall accuracy up to 2mm of target | ||||
| Optical Tracking system with mobile mounting solution | ||||
| TMS Coil Tracking tools including | ||||
| Calibration board and coil tracker available for all standard TMS coils | ||||
| Head reference that can attached to the patient’s head directly on skin by using self-adhesive medical patches or with clip onto a head band | ||||
| High precision pointer tool with rounded steel tip to navigate target 3D positions | ||||
| Dual-face coil tracker tools enable wider range of coil position and orientation | ||||
VMove
| Ingress Protection | : | IP54 (robot arm), IP32 (control box), IP40(robot stick) |
| Cleanroom Class | : | ISO Class 3 |
| Operating Temperature and Humidity | : | 0 to 50ËšC, 85% max. (with no condensation) |
| Operating and Storage Environment | : | No corrosive gases or exposure o water-soluble cutting oil. •1 |
| I/O Ports (Control Box) | : | Digital In: 16, Digtal Out: 16,Analog In: 2,Analog Out: 1 |
| I/O Ports (Tool) | : | Digital In: 4, Digtal Out: 4, Analog In: 1, Analog Out: 0 |
| I/O Interface | : | COM:3, HDMI: 1,LAN:3,USB2.0: 4, USB3.0: 2 |
| Communication | : | RS232, Ethernet (master),Modbus TCP/RTU (master & slave), PROFINET (optional),EtherNet/lP (optional) |
| I/O Power Supply | : | 24 VDC 2.0 A (control box) / 24 VDC 1.5 A (tool) |
| TMS Configuration: | ||
| Neuro-MSX main unit | : | 1 |
| Neuro-MSX cooling unit | : | 1 |
| Neuro-MSX Extra Power Supply Unit | : | 1 |
| Patient cap | : | 30 |
| Silicone oil canister | : | 1 |
| Angulated figure-of-eight cooled coil | : | 1 |
| Cooled ring coil | : | 1 |
| Double Cone Cooled Coil | : | 1 |
| Flexible arm for coil positioning | : | 1 |
| Coil holder (trolley/wall mounted) | : | 1 |
| Response Button | : | 1 |
| User manual «Neuro-MS.NET» soft copy | : | 1 |
| License for TMS | : | 1 |
| coil positioning tool | : | 1 |
| Navigation & Robotic: | ||
| Visor2 system | : | 1 |
| Visor2 essential accessories | : | 1 |
| Visor2 professional single lincense | : | 1 |
| Vmove robotic system | : | 1 |
| EEG Configuration: | ||
| Neuron-Spectrum-65 electronic unit | : | 1 |
| Stand for electronic unit | : | 1 |
| Stand mount | : | 1 |
| LED photic stimulator | : | 1 |
| Stand for LED photic stimulator | : | 1 |
| Reusable limb clamp ECG electrode | : | 4 |
| Cable for ECG channel | : | 1 |
| User manual "Neuron-Spectrum.NET" soft copy | : | 1 |
| EEG Cup electrode with cable | : | 30 |
| Electrode adhesive paste | : | 3 |
| Abrasive paste for skin preparation | : | 3 |
| Electrode system Caps | : | 1 |
| License for EEG | : | 1 |
| Equipment for Neurofeedback: | ||
| Inductive thorax and abdominal effort sensor | : | 1 |
| License for NeuroFeedback | : | 1 |
| Equipment for QEEG & Auditory Brain Stimulation: | ||
| Headphone | : | 1 |
| License for BASS software | : | 1 |
| License for QEEG software | : | 1 |
| Aksesoris Lokal: | ||
| PC atau Laptop (pilihan DELL atau setara) | : | 1 |
| UPS | : | 1 |
| Isolation Transformer | : | 1 |
| Printer (EPSON atau setara) | : | 1 |
| Treatment Chair Electric | : | 1 |
| Trolley | : | 1 |
| IP Camera | : | 1 |