| Stimulation Parameters |
|
|
| Pulse waveform |
: |
biphasic |
| Pulse duration range |
: |
240 … 410 µs |
| Peak magnetic field |
: |
2.5 T |
| Rated energy store on the main unit capacitor at the set amplitude of 100 % |
: |
320 J |
| |
|
|
| Stimulation Mode |
: |
Single Pulse |
| |
: |
Train |
| |
: |
Burt |
| |
: |
Ramp |
| |
: |
Sweep Frequency |
| Maximum frequency |
: |
100 Hz |
| Maximum pulse frequency in burst |
: |
2000 Hz |
| Coil surface temperature enabling automatic stimulation stop |
: |
(39 ± 2) °C |
| |
|
|
| Power Supply |
|
|
| Input voltage |
: |
(220 ± 22)/(230 ± 23)/(240 ± 24) V |
| Input frequency |
: |
50 or 60 Hz |
| Power consumption during the stimulation |
: |
|
| Main unit |
: |
≤ 3000 VA |
| Cooling unit |
: |
≤ 190 VA |
| Extra Power Supply unit |
: |
≤ 3000 VA |
| |
|
|
| Wireless Connection with PC |
|
|
| Interface |
: |
Wi-fi |
| |
: |
IEEE 802,11 b/g/n standard |
| General Specifications |
|
|
| Connection to PC |
: |
USB |
| Safety class |
: |
I |
| Applied partd |
: |
BF Type |
| Degree of protection provided by enclosure stimulator unit |
: |
IP20 |
| Operation temperature |
: |
+ 10 to + 35 ° C |
| Relative air humidity |
: |
20 to 85 % |
| |
|
|
| Stimulation Parameters |
|
|
| Pulse amplitue |
: |
0 - 100% |
| Adjustment step for pulse amplitude |
: |
1% |
| Intensity in % MT |
: |
0 - 150% |
| Adjustment step for intensity in % MT |
: |
1% |
| Adjustment stimulator discharge time |
: |
1 - 60 min |
| Adjustment step for automatic stimulator discharge time |
: |
1 min |
| |
|
|
| Train Mode |
|
|
| Pulse frequency in train |
: |
0.1 - 30 Hz |
| |
: |
0.1 - 100 Hz (option) |
| Permissible relative devition of pulse frequency in train |
: |
±5% |
| Adjustment step for pulse frequency in train |
|
|
| in the range 0.1 - 30 Hz |
: |
0.1 Hz |
| in the range 0.1 - 100 Hz |
: |
1 Hz |
| Number of pulse in train |
: |
1 - 2000 |
| Adjustment step for the number of pulse in train |
: |
1 |
| Maximum train duration |
: |
6000 s (calculation automatically depending on pulse frequency and number of pulse in train) |
| Intertrain interval |
: |
0 - 600 s |
| Permissible relative devition of intertrain interval |
: |
± 5% |
| Adjustment step for intertrain interval |
|
|
| in the range 0.1 - 30 Hz |
: |
0.1 s |
| in the range 0.1 - 100 Hz |
: |
1 s |
| Number of trains in session |
: |
1 - 1000 |
| Adjustment step for the number trains in session |
: |
1 |