Neuro MS/D with Neuronavigation & EMG

Description

  • Application: psychiatry, neurology
  • Solution for therapeutic repetitive transcranial magnetic stimulation (rTMS)
  • Number of pulses generated during one session – up to 10000
  • Neuro-MS.NET software to control magnetic stimulator

Specification

Stimulation Parameters    
Pulse duration    
Standard biphasic : 250–330 µs
Maximum frequency at maximum power : 5Hz²
With Extra Power Upply Unit : 20Hz²
     
Power Supply    
Supply voltage : (220±22)/(230±23) V
  : 50 or 60 Hz
Maximum power consumption during the stimulation: :  
Main unit : ≤1000 V·A
Cooling unit : ≤1000 V·A
Extra Power Supply Unit   ≤3000 V·A
Trig-in/out    
Trig-in : TTL/CMOS
Trig-out : TTL/CMOS
Trig-in pulse width : ≥ 200 µs
Stimulus delay relative to trig-in signal : 100...200 µs
Trig-out pulse width (user-defined) : 10 µs…50 ms

 

General Specifications and Parameters    
Interface : USB
Safety : I
Work parts : BF type
Cooling unit tank volume while filling up to the maximal level : 3.5 l
Type of cooling agent : Silicone oil 5 cSt
Ingress protection : IP20
Operation temperature : (10–35)º C
Relative air humidity : ≤80%
Cooling agent pressure in coil while cooling unit operating : not more than 0.15 MPa
Stimulation Parameters    
Stimulation frequency in repetitive stimulation mode    
Range of Permissible Values : (0.1–30) Hz (Stand-alone mode) & (0.1 – 100) Hz (PC Mode)
Frequency Step : 0.1 Hz (0.1 -10 Hz) & 1 Hz (10-100 Hz)
Stimulation frequency in burst* mode    
Range of Permissible Values : 50 – 100 Hz PC Mode
Frequency Step : 1 Hz
Pulse train duration range     
Range of Permissible Values : (0.5 - 100)s
Frequency Step : 0.5s(0 - 29.5s) & 0.1s(1.00 - 9.90s)
Inter-train interval range     
Range of Permissible Values : (0 - 300)s
Frequency Step : 0.5s(0 - 29.5s) & 0.1s(30 - 100s)
Session (0.5-40) min 0.5 min    
Range of Permissible Values : (0.5 - 40) min
Frequency Step : 0.5 min

 

Capacitor Charge Control    
Delay for the capacitor charge start (analysis epoch)**    
Range of Permissible Values : (0–9.90) s
Frequency Step : 0.01 s (0–0.99 s), 0.1 s (1.00–9.90s)
Trig-out    
Trig-out pulse width     
Range of Permissible Values : 10 μs, 1 ms–50 ms
Frequency Step : 1 ms

Standard Configurations

Neuro-MS/D main unit : 1
Neuro-MS/D cooling unit : 1
Neuro-MS/D Extra Power Supply unit : 1
End cap : 1
Patient cap : 20
Silicone oil canister : 1
Angulated figure-of-eight cooled coil : 1
Cooled ring coil : 1
Cooled Double Cone Coil : 1
Hydraulic system adapter : 2
HV Cable for extra power supply unit : 1
Control cable for Neuro-MS/D cooling unit : 1
Equipotential cable : 2
Mains supply cable : 3
USB сable : 1
Coil holder (trolley/wall mounted) : 2
flexible arm for coil positioning   1
User manual «Neuro-MS.NET» soft copy : 1
License for Neuro-MS.NET : 1
coil positioning tool : 1
Neurosoft equipment cover : 1
Silicon oil   1
Cable Trigger   1
     
EMG Configuration:    
Unit MEP : 1
Starter Kit EMG : 1
License for EMG : 1
     
Navigation Configuration:    
Neuronavigation Visor 2 System 1
Visor Essential accessories 1
Visor basic license set on USB dongle 1
     
Aksesoris Lokal:    
PC All in One : 1
Printer : 1
UPS : 1
Trolley TMS : 1
Comfort Treatment Chair : 1