For more than a quarter of a century, Neurosoft has been designing and producing various equipment for neurophysiology. Many years of experience and the meticulous work done by our software and hardware engineers have gone into the creation of the highest quality equipment, which is being successfully used at medical facilities all over the world.
Neuron-Spectrum-61 ensures:
|
|
|
EEG channels + differential channels for any physiological signal acquisition
Neuron-Spectrum-61 electronic unit features 11 EEG channels and 4 polygraphic channels for ECG, EOG, EMG and other signal type acquisition. The system also allows respiratory and SpO2 data recording. The channels can work both in AC and DC modes.
Perfect for cerebral function monitoring of neonates
Neuromonitor configuration of Neuron-Spectrum-61 is successfully used for cerebral function monitoring in the neonatal intensive care units (ICU). Simultaneous aEEG and native EEG recording allows for quick assessment of patient state which is crucial in ICU conditions and is indispensable when comprehensive EEG analysis is required.
Continuous impedance monitoring during acquisition
The color-coded impedance indication at the lead inputs and continuous impedance measurement during acquisition assist the specialist in signal quality monitoring and, if necessary, with timely adjustment of the electrode position, thus, ensuring superb recording quality. The operating mode can be switched at a push of a button on the front panel of the device whereas the LED indicator next to button will always show the current operating mode.
Biofeedback training with an optional software module
Biofeedback training is a non-pharmacological technique that helps patients learn to control physiological functions through neurofeedback to improve physical and mental health. This mind-body approach is used in patients with ADHD (attention deficit hyperactivity disorder), headache, addiction, anxiety, etc., as well as for muscle relaxation and developing of creative skills.
Neuron-Spectrum-61 with a special software module ensures various possibilities within BFB training:
|
|
|
|
Advanced tools for mathematical analysis of EEG
During EEG analysis the software automatically detects spikes and sharp waves indicative of abnormal EEG pattern. The excellent trace display quality is achieved by applying the cutting-edge smoothing technologies. The user-friendly solution for quick navigation throughout the entire EEG record streamlines the visual analysis.
The advanced tools for mathematical analysis of EEG are provided as well: amplitude, spectral and bispectral analysis, correlation and coherent analysis, wavelet-analysis, periodometric analysis, independent component analysis (ICA) are always available in all software configurations.
Standard electrode cap connector
Even if you use the electrode cap of various manufactures, it can be easily connected via the standard socket of an EEG amplifier. You no longer need any adapters or other accessories, which is time- and cost-effective.
|
|
|
|
|
|
| Amplifier | ||
| Number of Channel | : | 16 channels |
| 11 AC Channel | ||
| 1 ECG Channel | ||
| 4 Differential Channel | ||
| SpO2 (Optional) | ||
| Auditory Stimulator (Optional) | ||
| Sampling Rate | : | 16000 Hz |
| Voltage Range | : | 1 - 12000 μV |
| Ratio error Voltage Measurement | : | |
| in the range from 10 up to 50 μV | : | within ±25% |
| in the range from 50 up to 10000 μV | : | within ±7% |
| Sensitivity | : | 0.01 - 10.000.000 μV/mm |
| Ratio error of sensitivity | : | within ±5% |
| High pass filter | : | arbitary value can be set in 0.01 - 10 Hz range |
| Low pass filter | : | arbitary value can be set in 0.1 - 500 Hz range |
| Sweep Speed | : | arbitary value can be set in 1 - 10000 mm/s range |
| Ratio error sweep speed | : | within ±2% |
| Bandpass flatness in 0.5 - 60 Hz | : | from -10 up to +5% |
| Suppresion ratio of power frequency by notch filter | : | not less than 40 dB |
| CMRR | : | not less than 120 dB |
| Input noise level in 0.5 - 200 Hz range (rms) | : | not more than 1.4 μV |
| Input Impedance | : | not less than 400 MΩ |
| Patient leakage current | : | not more than 50 nA |
| Voltage measurement range | : | 0,2 - 100 mV |
| General Specifications and Parameters | ||
| Interface | : | USB |
| Supply voltage | ||
| electronic unit | : | 5 V DC |
| desktop PC based system | : | 220/230 V AC (50 Hz) |
| Electronic unit power consumption | : | not more than 2.8 VA |
| Safety | : | CF type |
| Minimum Personal Computer | ||
| i5, 6GB RAM, 500 GB HDD, Windows 10 | ||
| Color Printer & LED Monitor | ||
| Operating Limitations | ||
| Ambient temperature | : | 10 to 35 ะก |
| Relative humidity | : | 80 % at 25 C |
| Atmospheric pressure | : | 760±30 mmHg |
| Software | ||
| Export format avi, txt, rtf, xml, bmp, emf, EDF, pdf | ||
| Report automatically generated EEG report | ||
| EEG Software | ||
| Single / All channel editor | ||
| Bandpass Filter | : | Delta, Theta, Alpha, Beta, or custom rhytem |
| Spectral Analysis | : | Spectrum graph, spectrum table, spectrum map |
| Neurofeedback / Biofeedback | ||
| Brain Auditory Stimulator System (Delta, Tetha, Alpha, SMR & Gamma) |
| Neuron-Spectrum-61 electronic unit | : | 1 |
| Neurofeedback / Biofeedback Starter Kit | : | 1 |
| Electrode system Head Caps | : | 1 |
| User Manual Soft Copy | : | 1 |
| License for EEG software module | : | 1 |
| License for NFB/BFB | : | 1 |
| License for BASS software | : | 1 |
| Aksesoris Lokal | ||
| PC | : | 1 |
| Monitor | : | 2 |
| Printer | : | 1 |
| Trolley | : | 1 |
| UPS | : | 1 |
| Speaker | : | 1 |
| Headphone | : | 1 |