Neuro Audio (PTA)

Description

All-in-one Solution for ABR, ASSR, Multi-ASSR, MLR, LLR/CAEP, TEOAE, DPOAE, SOAE, ECochG, VEMP, P300, MMN and Other Tests

A full range of objective and subjective techniques for the diagnostics of acoustic and vestibular analyzer affection, hearing threshold search is in one small unit anow! Using this device you can perform ABR, ASSR, multi-ASSR, MLR, LLR/CAEP, TEOAE, DPOAE, SOAE, ECochG, VEMP, P300, MMN, pure tone audiometry techniques easily. 

Automatic Multi-ASSR Test

Multi-ASSR test implemented in Neuro-Audio system allows minimizing test duration due to automatization (intensity switching for hearing threshold search), stimulation of both ears simultaneously using 8 carrier frequencies (4 in each ear, from 500 Hz to 4000 Hz, ), use of frequency-specific Chirp stimulus which takes into account the cochlear anatomy for the response amplitude increase. Test result is represented as estimated pure tone audiogram used further for hearing aid fitting.

High Acquisition Quality

Well-considered system characteristics (low level of amplifier noise) and software features (adaptive notch filter, weighted averaging, Chirp stimulus, calculation of statistical parameters of signal and noise, up-to-date algorithms for automatic response detection) help a physician to obtain a high-quality record quickly. All tests can be performed when patient is awake (not-sedated).

VEMP with Biofeedback

Neuro-Audio system ensures full state-of-the-art support of VEMP tests, including VEMP with biofeedback, VEMP amplitude correction by mean EMG activity level, and also a unique function of EMG histogram with the possibility of post-hoc analysis (important for children). Acquisition of bone-conduction VEMP is also possible using bone vibrator amplifier (separate unit) and B-71 bone vibrator. Supported technique variants: cVEMP (cervical), oVEMP (ocular), gVEMP (galvanic).

New Standard for Routine Tests

Modern interface, powerful database, configurable normal values, high quality of obtained records and reproducible results allow user to solve a wide range of daily audiological problems with highest accuracy. Variety of built-in test templates helps you to start working with program, and the possibility of their adjustment makes it possible to set up all the templates for your own requirements. Due to these unique features Neuro-Audio system manufactured by Neurosoft can be used both in hospitals and research institutes.

Exellence

   
  • Air & Bone Conduction Testing
  • Flexible setting
  • New standard for routine tests
  • Customized reports
  • Reliable storage of patient records and test data
  • Contralateral Masking Noise (White Noise)
  • Automatic Mode (Hughson & Westlake), Manual Mode (with Mouse & Keyboard)
  • High-Frequency Audiometry (up to 16 KHz) with special headphones

Specification

           
  • Number of channels
: 2
  • Input impedance
: not less than 90 MΩ
  • Input voltage range
: 20 μV to 50 mV
  • Sampling rate
: 200 Hz – 80 kHz
  • The admissible error of voltage measurement should be within
:  
  • High pass filter
: 0.01–5000 Hz
  • in the voltage range from 20 to 100 μV
: ± 15 %
  • Low pass filter
: 10–10000 Hz
  • in the voltage range from 0.1до 50 mV
: ± 5 %
  • Notch filter
: not less than 40 dB
  • Gain
: 21-20832
  • Common-mode rejection
: not less than 100 dB
  • The admissible relative error of EP voltage measurement in the range from 1 to 400 μV at not less than 2000 averagings
: ± 10 %
  • Voltage range of the obtained averaged evoked potentials
: 0.1 to 400 μV
  • Input noise in 2–10000 Hz range (RMS)
: not more than 0.5 μV

 

   
           

Auditory stimulator :

  • Number of auditory channels
: 2
  • TDH-39 headband static force
  • Stimulation level (maximal for click)
: 137 dB SPL
  • Stimulation level (maximal for tone burst)
: Stimulation level (maximal for click)
  • Stimulation level (minimal for pure tone)
: -10 dB HL (or less)
  • Modulation frequency
: 20–200 Hz
  • Tone burst stimulation
:  
  • Continuous tone stimulation
: pure tone;
  • tone
: 20–10000 Hz
  • stimulus types
: amplitude, frequency, amplitude and frequency modulation
  • stimulus duration
: 0.05–50 ms
  • carrier frequency
: 125-8000 Hz
  • The sound pressure level at pure tone generation via the headphones
:  
  • Stimulus repetition rate
: 0.1–100 Hz with the admissible relative deviation of ± 0.1 %
  • at tone frequency within 125 – 4000 Hz range
: with ±3.0 dB accuracy
  • Click stimulation
:  
  • at tone frequency within 4 – 8 kHz range
: with ±5.0 dB accuracy
  • stimulus duration
: 50–5000 μs with the admissible relative deviation of ± 2 %
           

Maximum total harmonic distortion should not exceed :

  • for air conduction
: 2.5%
  • for bone conduction
: 5.5%
           

Chirp stimulus (cochlear delay compensation) :

  • wide-band
: 10–10000 Hz
  • narrow-band
: 499 Hz, 1 kHz, 2 kHz, 4 kHz
  • Stimulus polarity
: condensation, rarefaction, alternating
  • Noise masking intensity
  0–127 dB SPL (stimulus relative)
  • Frequency characteristics of masking noise
: White noise (0.02–10000 Hz)
  • Number of bone vibrator channels
: 0
  • Left/right/double-sided stimulation
: yes
  • Bone vibrator amplifier gain
: (19.5 ± 0.5) dB
  • Stimulus intensity at bone vibrator output
: -10 – 80 dB HL (at 3 kHz frequency)

 

   
           

General Parameters and Specifications :

  • Transport conditions
:  
  • Storage conditions
:  
  • temperature
: from - 50ºC up to +50ºC
  • temperature
: from +5ºC up to +40ºC
  • relative humidity
: 20%–95ºC
  • relative humidity
: 30%–85%
  • pressure
: from 70 kPa
  • pressure
: 70–106 kPa
  • Interface
: USB      

Standard Configurations

           
  • Neuro-Audio electronic unit

:

1
  • Auditory stimulator (headphones)
: 1
  • Patient button (with USB connector)
: 1
  • User manual "Neuro-Audio.NET"
: 1
  • USB hub with cable
: 1
  • License for the use of software "Neuro-Audio.NET" with additional software module "Neuro-Audio.NET/PTA"
: 1
           

Aksesoris Lokal:

  • PC (Dell atau setara)
: 1
  • Trolley
: 1
  • Printer (EPSON atau setara)
: 1
  • UPS
: 1